Technical specifications
System 2nd Generation SWA: modular, analytical system platform, consolidated work area for Clinical Chemistry and Immunology, expandable and re-configurable on site
System components
Control unit: PC, monitor, keyboard, printer, etc. on an ergonomic stand
Core unit: Sample input/output, sample transport by Intelligent Process Management
Analytical modules: 2 analytical modules
Type of module
cobas c 501 module: Photometric measuring unit (incl. ISE)
cobas e 601 module: ECL technology measuring unit
Number of module combinations
7 module combinations
up to 3 modules in one core unit
cobas c 501 modules: 1 to 2 modules
cobas e 601 modules: 1 to 2 modules
cobas c 501 | 601 modules: 2 to 3 modules (only 2 units of one module type for SWA combinations possible)
Sample throughput up to 120 racks/hr
or samples: 600/hr (theoretical max)
Test throughput (theoretical max) 170 up to 2,170 tests/hr - overall
1,000 up to 2,000 tests/hr - with cobas c 501 modules
170 up to 340 tests/hr - with cobas e 601 modules
1,170 up to 2,170 tests/hr - with photometric and HetIA modules
Number of channels (reagent slots) 25 up to 145 channels/reagent slots
3 ISE
60 channels (reagent slots, automatic setting) on cobas c 501 module
25 channels (reagent slots) on cobas e 601 module
Programmable parameters 191 parameters with photometric and HetIA modules
117 photometric tests, 3 ISE tests, 8 formulas, 3 serum indices with photometric modules
60 heterogeneous tests with cobas e 601 modules
Sample material Serum, Plasma, Urine, CSF
Sample input/output on
Core unit analytics
Load/unload capacity: 150 samples/30 racks on 2 trays
Rack: 5 position rack, RD standard rack
Rack types: Routine, STAT, Control, Calibrator, Rerun (manual), Wash
Tray: Tray with 15 racks/75 samples, RD standard tray
STAT port: STAT samples are processed with priority
Sample container types
Primary tubes: 5 to 10 ml; 16 x 100, 16 x 75, 13 x 100, 13 x 75 mm
Sample cup: 2.5 ml
Micro cup: 1.5 ml, (exception: cobas e 601 module)
Cups on tube: Cup on top of a 16 x 75/100 mm tube
Cup on top of one non standard tube
False bottom tube: one type is definable
Sample volume 1 - 35 µl
Sample dilution 3 - 121 times, diluent > 100 µl
Sample clot detection Available for cobas c 501 and cobas e 601 module
Minimum sample volume
Primary tubes: 700 µl
Sample cup: 100 µl
Micro cup: 50 µl
Sample barcode types Code 128
Codabar (NW 7)
Interleaved 2 of 5
Code 39
Control unit

PC: Windows XP, Pentium IV processor, DVD -RAM
Monitor: 17’ TFT touch screen colour monitor
Keyboard 101 – key enhanced, country-specific
Printer: optional
Pc stand: optional, ergonomic (UL, CE, GS, TüV)
e- link for cobas® modular plattform

System interfaces RS 232 serial interface, bi-directional
Interface to e-services for the cobas® modular plattform (teleservices functionalities and automatic download)
Sample data base 10,000 routine / STAT samples
Test methods For photometric modules: 1 point, 1 point + prozone check, 2 point, 2 point kinetic, 2 point +prozone check, 3 point, 1 point + kinetics, Rate A, Rate A + serum index, Rate A with blank, Rate B
Calibrator/QC input Via specified racks through the input buffer or STAT port
Calibration methods Start-up, Re-calibration
For photometric modules: Linear, non-linear multi-points, 2 point calibration, K-factor up to 100 different calibrators pre programmable
Storage of up to 180 curves
preventive calibration of the stand-by cobas c pack and cobas e pack
on cobas c 501: Two k-factor can be defined for different sample types
QC methods For e- and photometric modules: Real-time QC, individual QC, cumulative QC up to 100 controls pre programmable
preventive QC after calibration of the stand-by cassettes/ rack packs
Auto QC: QC without operator intervention (timer triggered)
Programmable parameters 191 parameters with photometric and HetIA modules
117 photometric tests, 3 ISE tests, 8 formulas, 3 serum indices with photometric modules
60 heterogeneous tests with cobas e 601 modules
Rerun/reflex function Automatic rerun and manual rerun
Automatic reflex is supported by the system, reflex request to be provided by PSM or LIS
Electrical requirements
Power requirements: 230 Volts AC
110 Volts AC
4.5 kVA (for cobas® 6000 < 501 | 601 >)
Frequency: 50 Hz or 60 Hz +/- 0.5 %
Water/waste requirements
Water: Bacteria free, deionised water supply: resistance of < 1,0 µS/cm
Water consumption: under routine conditions in average 15 l/hr , max. 30 l/hr (per module)
Water pressure: 0.5-3.5 kg / cm3, (49-343 Kpa)
Biohazards waste: Separate container behind the system
L. concentrated
wall drain < waste:

Central drain port, diameter: >= 50 mm
100 mm above the floor
Sample container types
Primary tubes: 5 to 10 ml; 16 x 100, 16 x 75, 13 x 100, 13 x 75 mm
Sample cup: 2.5 ml
Micro cup: 1.5 ml, (exception: cobas e 601 module)
Cups on tube: Cup on top of a 16 x 75/100 mm tube
Cup on top of one non standard tube
False bottom tube: one type is definable
Regulatory requirements GS, CE, UL, C-UL
Operating conditions
Ambient temperature: 15 to 32 °C
Ambient humidity: 45 to 85 % (RH, without condensation)
Heat output: 1.5 kW (5,400 kJ/hr for cobas c 501, 4,320 kJ/hr for cobas e 601)
Noise output: < 68 dB
Physical dimensions
Width: 1,850 to 4,250 mm
Core unit: 60 cm; per cobas c or cobas e module: 120 cm
Depth: 98 cm; per cobas c or cobas e module
Height: 130 cm
Control unit: Width: 80 cm, depth: 80 cm
Weight
425 to 1,000 kg  
cobas c 501 module: approx. 350 kg
cobas e 601 module: approx. 380 kg
Core unit: approx. 217 kg
Control unit: approx. 75 kg